Página 1 - Handheld Digital Multimeter

Handheld Digital Multimeter GDM-360, GDM-397, GDM-398, GDM-461 USER MANUAL GW INSTEK PART NO. 82DM-46100MA1 ISO-9001 CERTIFIED MANUFACTURER

Página 2

OVERVIEW 8  Use the proper terminals, function, and range for your measurements.  Do not use or store the Meter in an environment with high temp

Página 3 - Table of Contents

GDM-360/397/398/461 User Manual 9  A soft cloth and mild detergent should be used to clean the surface of the Meter when servicing. No abrasives

Página 4

OVERVIEW 10 The Meter Structure (see figure 1) 1. LCD display 123456 2. Functional buttons 3. Yellow button 4. Blue button 5. Rotary switch 6.

Página 5 - Table of Contents

GDM-360/397/398/461 User Manual 11 Rotary Switch The table below describes the relevant functions of each of the rotary switch positions. Rotary Sw

Página 6 - OVERVIEW

OVERVIEW 12 Functional Buttons The table below describes what operations are performed for each of the buttons. Button Operation Performed LIGHT (ex

Página 7 - Unpacking Inspection

GDM-360/397/398/461 User Manual 13 *Note: For the GDM-398, holding the REL key 2 seconds will also display the DATA OUTPUT symbol, however the

Página 8 - Safety Information

OVERVIEW 14 11 The continuity buzzer is on. 12 MAX/MIN Maximum and Minimum reading. 13 Data output is in progress. Note: This symbol has no funct

Página 9 - Rules for Safe Operation

GDM-360/397/398/461 User Manual 15 mF: Milli-farad. 1 x 10-3 or 0.001 farads µF: Microfarad. 1 x 10-6 or 0.000001 farads. nF: Nanofarad. 1 x 10-9 o

Página 10 - OVERVIEW

MEASUREMENT OPERATION 16 MEASUREMENT OPERATION A. DC/AC Voltage Measurement (see figure 2) V GDM-360/397/398 AV GDM-461 figure 2 Note The GDM-360 d

Página 11

GDM-360/397/398/461 User Manual 17 Warning To avoid injury to your person or damage to the Meter from electric shock, please do not attempt to meas

Página 12 - The Meter Structure

This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied,

Página 13 - Rotary Switch

MEASUREMENT OPERATION 18 Note  In each range, the Meter has an input impedance of 10MΩ except for the mV range which has an input impedance of 300

Página 14 - Functional Buttons

GDM-360/397/398/461 User Manual 19 B. DC/AC Current Measurement (see figure 3) A GDM-360/397/398 A GDM-461 figure 3

Página 15 - Display Symbols

MEASUREMENT OPERATION 20 Warning Before connecting the Meter to the return circuit to be tested, cut off the current of the return circuit. If the

Página 16

GDM-360/397/398/461 User Manual 21 GDM-360/461: Displays the true rms value. 5. Press Hz% (GDM-360/397/398) or the yellow button (GDM-461) to obt

Página 17

MEASUREMENT OPERATION 22 C. Resistance Measurement (see figure 4) mAuA COM VΩHz figure 4

Página 18 - MEASUREMENT OPERATION

GDM-360/397/398/461 User Manual 23 Warning To avoid damaging the Meter or the DUT, disconnect power from the circuit and discharge all high-voltage

Página 19

MEASUREMENT OPERATION 24 Note  The test leads can add 0.2Ω to 0.5Ω of error to the resistance measurement. To obtain precise readings for low-resi

Página 20 - MEASUREMENT OPERATION

GDM-360/397/398/461 User Manual 25 D. Testing for Continuity (see figure 5) figure 5

Página 21 - B. DC/AC Current Measurement

MEASUREMENT OPERATION 26 Warning To avoid damage to the Meter or to the device under test, disconnect the circuit power and discharge all high-volta

Página 22

GDM-360/397/398/461 User Manual 27 E. Testing Diodes (see figure 6) figure 6

Página 23

Table of Contents 1 Table of Contents OVERVIEW...

Página 24 - C. Resistance Measurement

MEASUREMENT OPERATION 28 Warning To avoid possible damage to the Meter and to the device under test, disconnect the circuit power and discharge all

Página 25

GDM-360/397/398/461 User Manual 29  Connect the test leads to the proper terminals as said above to avoid errors. The LCD will display OL to ind

Página 26

MEASUREMENT OPERATION 30 F. Capacitance Measurement (see figure 7) mAuA VΩHz figure 7

Página 27 - D. Testing for Continuity

GDM-360/397/398/461 User Manual 31 Warning To avoid damage to the Meter or to the equipment under test, disconnect the circuit power and discharge

Página 28

MEASUREMENT OPERATION 32 3. Connect the test leads to the DUT. The measured value shows on the display. Note  It takes a longer time to measure a

Página 29 - E. Testing Diodes

GDM-360/397/398/461 User Manual 33 G. Frequency Measurement (see figure 8) GDM-360/397/398 GDM-461 figure 8 Warning To avoid personal harm, do n

Página 30

MEASUREMENT OPERATION 34 2. Set the rotary switch to Hz%; frequency measurement is set to (Hz) as default, otherwise press the Hz% button (GDM-360/

Página 31

GDM-360/397/398/461 User Manual 35 H. Temperature Measurement (GDM-397 only) (see figure 9) figure 9

Página 32 - F. Capacitance Measurement

MEASUREMENT OPERATION 36 Warning To avoid harm to yourself, do not input higher than DC 60V or AC 30V voltages. To measure temperature, connect th

Página 33

GDM-360/397/398/461 User Manual 37 I. Transistor hFE Measurement (GDM-398 only) (see figure 10) figure 10

Página 34

GDM-360/397/398/461 User Manual 2 MAX MIN button (GDM-360/397/398 only) ...

Página 35 - G. Frequency Measurement

MEASUREMENT OPERATION 38 1. Set the rotary switch to hFE. 2. Insert the multi-purpose socket into the input terminal as shown on figure 10. 3. In

Página 36

GDM-360/397/398/461 User Manual 39 J. EF Function (GDM-398 only) (see figure 11) figure 11

Página 37 - (see figure 9)

MEASUREMENT OPERATION 40 To use the EF function, connect the Meter as follows: 1. Set the rotary switch to EF and remove the test lead from the inp

Página 38

GDM-360/397/398/461 User Manual 41 Hold Mode Warning To avoid possibility of electric shock, do not use Hold mode to determine if circuits are with

Página 39 - (see figure 10)

MEASUREMENT OPERATION 42 RANGE button  Press RANGE to enter the manual ranging mode; the Meter beeps.  Press RANGE to step through the ranges av

Página 40

GDM-360/397/398/461 User Manual 43  Press PEAK to step the display through Peak Max and Peak Min readings.  Press and hold PEAK for over 2 seco

Página 41 - (see figure 11)

MEASUREMENT OPERATION 44 The Use of Relative Value Mode The REL mode applies to all measurement functions except frequency/duty cycle measurement.

Página 42

GDM-360/397/398/461 User Manual 45 The BLUE button It is used for selecting the required measurement function when there is more than one function

Página 43 - Hold Mode

GENERAL SPECIFICATIONS 46 GENERAL SPECIFICATIONS  Maximum Voltage between any Terminals and Ground: Refer to the different input ranges for the

Página 44 - Peak Hold (GDM-461 only)

GDM-360/397/398/461 User Manual 47  Overload indication: Display OL  Battery Deficiency: Display  Temperature:  Operating: 0˚C to

Página 45

Table of Contents 3 Replacing the Fuses ...

Página 46

GENERAL SPECIFICATIONS 48 Accuracy Specifications Accuracy: ±(a% reading + b digits), guaranteed for 1 year. Operating temperature:18˚C ~28˚C. Relat

Página 47 - The BLUE button

GDM-360/397/398/461 User Manual 49 GDM-360 Range Resolution Accuracy Input Impedance Fixed Value Input 60mV 0.01mV ±(0.8%+3) Around >3000MΩ 1000

Página 48 - GENERAL SPECIFICATIONS

GENERAL SPECIFICATIONS 50 AC Voltage GDM-397/398 Range Resolution Accuracy 45~400Hz Input Impedance Fixed Value Input GDM-398 GDM-397 40mV 0.01mV

Página 49

GDM-360/397/398/461 User Manual 51 GDM-360 Range Resolution Accuracy Input Impedance Fixed Value Input 45~1kHz >1kHz~3kHz 60mV 0.01mV ±(1.2%+5)

Página 50 - DC Voltage

GENERAL SPECIFICATIONS 52 GDM-461 Range Resolution Accuracy Input Impedance Fixed Value Input 45~1kHz >1kHz~10kHz 220mV 0.01mV ±(1.0%+10) ±(1.5%+

Página 51 - GDM-461

GDM-360/397/398/461 User Manual 53 DC Current GDM-397/398 Range Resolution Accuracy Overload Protection 400μA 0.1μA ±(1.0%+2) Fuse 1: F600mA H 1000

Página 52 - AC Voltage

GENERAL SPECIFICATIONS 54 GDM-360 Range Resolution Accuracy Overload Protection 600μA 0.1μA ±(1.0%+3) Fuse 1: F600mA H 1000V,φ6.35 x 31.8mm 6000μA 1

Página 53 -  GDM-360:

GDM-360/397/398/461 User Manual 55 GDM-461 Range Resolution Accuracy Overload Protection 220μA 0.01μA ±(0.5%+10) Fuse 1: F600mA H 1000V,φ6.35 x 31.

Página 54 - GENERAL SPECIFICATIONS

GENERAL SPECIFICATIONS 56 AC Current GDM-397/398 Range Resolution Accuracy 45Hz~400Hz Overload Protection 400μA 0.1μA ±(1.2%+5) Fuse 1: F600mA H 10

Página 55 - DC Current

GDM-360/397/398/461 User Manual 57 GDM-360 Range Resolution Accuracy Overload Protection 45~1kHz 1kHz~3kHz 600μA 0.1μA ±(1.2%+5) ±(1.5%+5) Fuse 1:

Página 56

OVERVIEW 4 OVERVIEW This User Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the W

Página 57

GENERAL SPECIFICATIONS 58 GDM-461 Range Resolution Accuracy Overload Protection 45~1kHz >1kHz~10kHz 220μA 0.01μA ±(0.8%+10) ±(1.2%+50) Fuse 1: F6

Página 58 - AC Current

GDM-360/397/398/461 User Manual 59 Resistance GDM-397/398 Range Resolution Accuracy Overload Protection Remark 400Ω 0.1Ω ±(1.2%+2) 1000V dc / ac

Página 59

GENERAL SPECIFICATIONS 60 GDM-360 Range Resolution Accuracy Overload Protection Remark 600Ω 0.1Ω ±(1.2%+2) 1000V dc / ac (Vpp) When measuring below

Página 60

GDM-360/397/398/461 User Manual 61 GDM-461 Range Resolution Accuracy Overload Protection Remark 220Ω 0.01Ω ±(0.5%+10) 1000V dc / ac (Vpp) When mea

Página 61 - Resistance

GENERAL SPECIFICATIONS 62 Capacitance GDM-397/398 Range Resolution Accuracy Overload Protection Remark 40nF 0.01nF ±(3.0%+5) 1000V dc / ac (Vpp) T

Página 62

GDM-360/397/398/461 User Manual 63 GDM-360 Range Resolution Accuracy Overload Protection Remark 40nF 0.01nF ±(3.0%+5) 1000V dc / ac (Vpp) There is

Página 63

GENERAL SPECIFICATIONS 64 GDM-461 Range Resolution Accuracy Overload Protection Remark 22nF 0.001nF ±(3.0%+5) 1000V dc / ac (Vpp) There is around 5

Página 64 - Capacitance

GDM-360/397/398/461 User Manual 65 Frequency Model Range Accuracy Maximum Resolution GDM-397/398/360 10Hz~10MHz ±(0.1%+4) 0.01Hz GDM-461 10Hz~220MH

Página 65

GENERAL SPECIFICATIONS 66 Diode Test Model Resolution Remarks Overload Protection GDM-360/397/398 0.001V Open circuit voltage around 2.8V 1000Vdc /

Página 66

GDM-360/397/398/461 User Manual 67 Temperature Measurement (GDM-397 only) Range Resolution Accuracy Overload Protection ˚C 1˚C (-40~-20˚C): -(8%+5)

Página 67 - Frequency

GDM-360/397/398/461 User Manual 5 Unpacking Inspection Open the package case and take out the Meter. Check the following items carefully to see if

Página 68 - Continuity Test

MAINTENANCE 68 MAINTENANCE This section provides basic maintenance information including battery and fuse replacement instructions. Warning Do not

Página 69 -  Thermocouple:

GDM-360/397/398/461 User Manual 69  Do not use or store the Meter in a place of humidity, high temperature or in the presence of explosives, infl

Página 70 - MAINTENANCE

MAINTENANCE 70 Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the

Página 71 - Replacing the Battery

GDM-360/397/398/461 User Manual 71 Replacing the Fuses figure 13

Página 72 - MAINTENANCE

MAINTENANCE 72 Warning To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified fuses ONLY in accordance wit

Página 73 - Replacing the Fuses

GDM-360/397/398/461 User Manual 73 10A range: F2, 10A H 1000V,φ10.3x38.1mm. (CE) 6. Rejoin the case bottom and case top, and reinstall the scre

Página 74

OVERVIEW 6 Safety Information This Meter complies with the EN61010-1 and EN61010-2-030 standards: in pollution degree 2, overvoltage category (CAT.

Página 75 - USB and RS232C Serial Port

GDM-360/397/398/461 User Manual 7 Rules for Safe Operation Warning To avoid possible electric shock or personal injury, and to avoid possible damag