DC Electronic Load PEL-3000 Series USER MANUAL VERSION: 1.14 ISO-9001 CERTIFIED MANUFACTURER

Página 2

PEL-3000 User Manual 8 GETTING STARTED This chapter provides a brief overview of the PEL-3000, the package contents, instructions for first time u

Página 3 - Table of Contents

PEL-3000 User Manual 98 4. Repeat step 3 for any remaining programs in the chain. 5. Press Save to save the program chain to internal memory.

Página 4

OPERATION 99 5. When a program/chain has finished running, a list of the Go-NoGo results for each step are displayed.  Press Exit[F5] to exit. D

Página 5 - SAFETY INSTRUCTIONS

PEL-3000 User Manual 100 Sequence The PEL-3000 supports both programs and sequences. The essential difference between programs and sequences is t

Página 6 - Safety Guidelines

OPERATION 101 Sequence 1STEP 01STEP 02:STEP NSequence 2STEP 01STEP 02:STEP NSequence 10STEP 01STEP 02:STEP N Start Sequence: ::: Description Norm

Página 7

PEL-3000 User Manual 102 Memo 12 characters A user-created note for the currently selected sequence. Mode CC, CR, CV, CP Operating mode for the se

Página 8

OPERATION 103 RAMP ON, OFF When turned on the current transition is evenly ramped from the start of the step to the end of the step. When turned off

Página 9

PEL-3000 User Manual 104 PAUSE ON, OFF Pause: Inserts a pause at the end of the step. When paused, the unit will pause at the end of the step curr

Página 10 - GETTING STARTED

OPERATION 105 4. Set the following parameters for the currently selected sequence. See page 100 for details on each parameter.  Memo  Mode 

Página 11

PEL-3000 User Manual 106 Data Edit Configuration Data Edit Display PreviousMenuLOADRS23203/Sep/2012Data Edit for Normal SequencesRAMP:LOAD:OFFOF

Página 12 - PEL-3000 Series Introduction

OPERATION 107 5. Set the following parameters for the currently selected step. See the Data Edit Overview on page 102 for configuration details. 

Página 13 - Main Features

GETTING STARTED 9 (PEL-3211 Booster Pack) ... 21 Display ...

Página 14 - Accessories

PEL-3000 User Manual 108 Running a Normal Sequence Description A load created with the Normal Sequence function is run the same way as a normal l

Página 15

OPERATION 109 Display: Sequence/Chain Running ContinueNSEQRS23203/Sep/2012Run N.Seq.0.000V0.000A0.00wStepSeq. No: 01Loop: 00010003

Página 16 - Package Contents

PEL-3000 User Manual 110 Fast Sequence Overview Description A fast sequence is comprised of a user-defined number of steps that can be executed a

Página 17 - Appearance

OPERATION 111 Description Fast Sequence configuration is split into Timing Edit configuration and Data Edit configuration. Timing Edit configuratio

Página 18 - (PEL-3211 Booster Pack)

PEL-3000 User Manual 112 Data Edit Overview Each step in a fast sequence contains the following setting parameters: Setting Setting Range De

Página 19

OPERATION 113 range. Value settingStep 02Step 4Step 01StepFILL exampleStep 03End_ValueStart_Value Filled values Filled steps Start_Step End_Step

Página 20

PEL-3000 User Manual 114 Timing Edit Configuration Edit Timing Display PreviousMenuLOADRS23203/Sep/2012Timing Edit for Fast SequenceLoop:Range:M

Página 21 - TRIG OUT

OPERATION 115 Sequence Timing configuration is complete.  Go to Data Edit to edit the steps used in the Fast Sequence. Page 115.  Go to Runni

Página 22 - (PEL-3111)

PEL-3000 User Manual 116 2. Press Insert Point[F1] to add a step to the sequence.  Every-time Insert Point is pressed the Step parameter is in

Página 23 - FRAME CONT

OPERATION 117 Save After all the steps for the sequence are complete, press Save[F3] to save the steps. Data Edit for Fast Sequences configuration

Página 24

PEL-3000 User Manual 10 PEL-3000 Series Introduction The PEL-3000 Series is a family of high performance DC electronic loads positioned to test a

Página 25 - Display

PEL-3000 User Manual 118 Display: Fast Sequence Running FSEQRS23203/Sep/2012Run F.Seq.0.998V153.10mA0.1528wStepLoop: 00010023Current loop numberCu

Página 26 - First Time Use Instructions

OPERATION 119 Save Recall The PEL-3000 can save and recall system settings, preset data, memory data, Go-NoGo settings as well as normal and fast s

Página 27

PEL-3000 User Manual 120 For example: To load Preset Data P7 from USB, you must first load Preset Data P0~P9 to internal memory, then from inter

Página 28

OPERATION 121 Preset Data Preset Data contains the same settings as the Memory Data. Preset Data contains the operating mode, range, response and Go

Página 29 - Power Up and Self Test

PEL-3000 User Manual 122 Memory Data Example Active settingM001::M256 MXXXMedia | Memory Display SaveData Type MemoryRS23203/Sep/2012Memory M256A

Página 30 - Setting the Date and Time

OPERATION 123 Note Normal Sequence and Fast Sequence data cannot be recalled from or saved to an internal memory slot. Saving Files to USB Memory

Página 31 - Load Wiring

PEL-3000 User Manual 124 Operation 1. Insert a USB drive into the USB port. 2. Press Shift + FileFUNC. 3. Select USB with the Media[F1] soft-ke

Página 32

OPERATION 125 Recalling Files from Internal Memory Description When recalling Memory, Setup or Preset Data from the internal memory slots, the reca

Página 33 - Twisted

PEL-3000 User Manual 126 4. Select which memory slot to recall from. Memory: Setup Memory: Preset: M001 ~ M256 1 ~ 100 P0 ~ P9 5. Press Recal

Página 34 - Load Wire Connections

OPERATION 127 Memory Data Example Media | USBMedia | MemoryRecall fileM001::M256 MXXX For example, if the file 3021_01.M is recalled, all the Memor

Página 35

GETTING STARTED 11 Main Features Performance  High slew rates of up to 16A/μS(PEL-3111) for a fast response speed  High capacity when used in p

Página 36 - + potential

PEL-3000 User Manual 128 Memory: Setup Memory: Preset: NSeq: FSeq: Model_file number.M Model_file number.S Model_file number.P Model_file number.

Página 37 - Using the Terminal Cover

OPERATION 129 Note This setting only applies when recalling preset settings from internal memory, either by using the Presets keys (P0 ~ P9) or by

Página 38

PEL-3000 User Manual 130 Rename a Folder 1. Use the scroll wheel to move the cursor to the file/folder you wish to rename. 2. Press Rename[F3].

Página 39 - GETTING STARTED

OPERATION 131 Quick Preset Recall Description Presets P0 to P9 can be recalled quickly by using the Preset key and the number pad. Operation 1. Pr

Página 40 - Remote Sense

PEL-3000 User Manual 132 User’s Default Setting Description The currently active settings can be set as the ―User’s Default‖ settings. Save User’

Página 41 - Firmware Update

EXTERNAL CONTROL 133 EXTERNAL CONTROL Analog Control ... 134 J1 Connector Overview ...

Página 42 - Serial Number : 00000001

PEL-3000 User Manual 134 Analog Control The Analog Control subsection describes how to use the J1 Frame Control Connector for voltage or resistanc

Página 43 - Conventions

EXTERNAL CONTROL 135 External Voltage Control - Overview Background External voltage control of the CC, CR, CV and CP mode is accomplished using th

Página 44 - Parameter selection

PEL-3000 User Manual 136 Caution When using external voltage control, make sure no more than ±11V is applied across pins 1 and 3. Exceeding this

Página 45 - Enter key

EXTERNAL CONTROL 137 CR Mode Input conductance = rated conductance × (external voltage/10) Input ConductanceExternal VoltageRated Conductance0V 10V

Página 46 - Parameter

PEL-3000 User Manual 12 Accessories Standard Accessories Part number Description 82EL-31110MB1 Quick Start Guide 82EL-31110EB1 User / Program

Página 47

PEL-3000 User Manual 138 3. Turn the power on the PEL-3000. 4. Set the operating mode and range.  See page 51 for CC mode.  See page 53 fo

Página 48 - Character

EXTERNAL CONTROL 139 Connection When connecting the external resistance source to the J1 connector, use a ferrite core and use twisted pair wiring.

Página 49 - Help Menu

PEL-3000 User Manual 140 External Resistance Control – Operation Description External resistance control can be used to control the current, volt

Página 50 - OPERATION

EXTERNAL CONTROL 141 CR Mode Proportional Control: Input conductance = rated conductance × (external resistance/10). Inverse Control: Input conduct

Página 51

PEL-3000 User Manual 142 CP Mode Proportional Control: Input power = rated power × (external resistance/10). Inverse Control: Input power = rated

Página 52

EXTERNAL CONTROL 143 5. Press Main > Configure [F5] > Next Menu [F4] > External [F3]. 6. Set the Control to R for proportional control or

Página 53 - Basic Operation

PEL-3000 User Manual 144 Example The LoadOn IN setting determines whether the load is turned on when the external switch is closed (low) or open (

Página 54

EXTERNAL CONTROL 145 Load On/Off Status Description Pin 13 (Load On Status) of the J1 connector is used to monitor the load status (on or off). P

Página 55

PEL-3000 User Manual 146 Pin Inputs Pins 8 and 9 of the J1 connector are internally pulled up to 5V with a 10kΩ resistor when open. When closed, p

Página 56

EXTERNAL CONTROL 147 Pin out The Range Status pins are photo-coupled open-collector outputs. 1714, 15 Photocoupler input: 30V max, 8mA, max. Exter

Página 57

GETTING STARTED 13 GTL-255 300mm Frame Link Cable (for linking units that are stacked) GTL-248 GPIB cable, 2.0m GTL-246 USB cable, Type A - Type

Página 58

PEL-3000 User Manual 148 External Control of the Alarm Description An alarm can be activated/deactivated using external control with the J1 conne

Página 59

EXTERNAL CONTROL 149 Alarm Status Description Pins 16 and 17 of the J1 connector are used to monitor whether the alarm is on or off. Pin out The A

Página 60 - +CV Mode

PEL-3000 User Manual 150 Monitor Signal Output Trigger Signal Output Description The trigger output signal is generated every time a switching op

Página 61 - Turning on the Load

EXTERNAL CONTROL 151 settings. CurrentIMON VRated Current Monitor Output Range Monitor Connector Current Range Monitor Output Range I MON OUT (B

Página 62 - Shorting the Load

PEL-3000 User Manual 152 Parallel Operation The PEL-3000 series can be connected in parallel to increase the total power capacity of a single unit

Página 63 - Short Key Configuration

EXTERNAL CONTROL 153 Parallel Capacity, PEL-3211 Model No. of Units V I Total Sink Current PEL-3111 + PEL-3211 Total Power PEL-3111 + PEL-3211 P

Página 64 - LOCK icon

PEL-3000 User Manual 154 MASTER+-+S-SSLAVE 1+-+S-S+-+S-SSLAVE 2+-+S-SDUT-+J2J1J2J1J2J1J2J1SLAVE 3Load wiringFrame link cable(GTL-255)Sense wiring

Página 65 - Basic Configuration

EXTERNAL CONTROL 155 Configuration Description When using the multiple units in parallel all the basic settings are adopted from the master unit. O

Página 66

PEL-3000 User Manual 156  A maximum of 5 units can be used in parallel.  A maximum of 4 boosters can be used with a single PEL-3111, acting as

Página 67 - +CV OFF V

EXTERNAL CONTROL 157 Turning the Load On Description Operating the PEL-3000 Series in parallel mode is the same as for single units. Note When u

Página 68 - % from the Set value

PEL-3000 User Manual 14 Package Contents Check the contents before using the instrument. Opening the box Contents (single unit)  Main

Página 69

PEL-3000 User Manual 158 Disable Parallel Mode Description To disable parallel mode, each unit must be set as a ―Master‖. Operation 1. Turn the

Página 70 - CV Mode Response Speed

REMOTE CONTROL 159 REMOTE CONTROL This chapter describes basic configuration of IEEE488.2 based remote control. For a command list, refer to the pro

Página 71 - CV A Value 15.000

PEL-3000 User Manual 160 Interface Configuration Configure to USB Remote Interface USB configuration PC side connector Type A, host PEL-3000 side

Página 72 - Soft Start Setting

REMOTE CONTROL 161 2. Connect a GPIB cable from a GPIB controller to the GPIB port on the PEL-3000. 3. Turn the PEL-3000 on. 4. Press Shift + Ut

Página 73 - Von Voltage Settings

PEL-3000 User Manual 162 Configure RS232C RS232C Configuration Connector DB-9, Male Baud Rate 2400, 4800, 9600, 19200, 38400 Stop Bit 1, 2 Parit

Página 74 - Von Voltage Latch

REMOTE CONTROL 163 RS232C/USB Remote Control Function Check Functionality check Invoke a terminal application such as Realterm. For RS-232C, set t

Página 75 - Timer Functions

PEL-3000 User Manual 164 Using Realterm to Establish a Remote Connection Background Realterm is a terminal program that can be used to communic

Página 76 - Cut Off Time

REMOTE CONTROL 165 If using USB, the baud rate, stop bit and parity settings can be viewed by right-clicking connected device and selecting the Prop

Página 77 - Auto Load Configuration

PEL-3000 User Manual 166 6. After Realterm has started, click on the Port tab. Enter the Baud, Parity, Data bits, Stop bits and Port number conf

Página 78

REMOTE CONTROL 167 7. Click on the Send tab. In the EOL configuration, check on the +CR and +LF check boxes. Enter the query: *idn? Click on Send

Página 79 - Cursor Mode Configuration

GETTING STARTED 15 Appearance PEL-3000 Front Panel (PEL-3021/PEL-3041) I MON OUT TRIG OUT1.5 - 150V175W0 - 35AP0P1P4P7CAL.P2P5P8LockP3P6P9UtilityL

Página 80 - Step Mode Configuration

PEL-3000 User Manual 168 GPIB Function Check Functionality check Please use the National Instruments Measurement & Automation Controller so

Página 81

REMOTE CONTROL 169 3. Press the Scan for Instruments button. 4. In the Connected Instruments panel the PEL-3000 should be detected as Instrument

Página 82 - Protection Settings

PEL-3000 User Manual 170 678 9. The function check is complete.

Página 83 - Alarm message

FAQ 171 FAQ • The load voltage indicated on the load module is below expected. • The front panel keys are not working. • The load won’t turn on. •

Página 84

PEL-3000 User Manual 172 For more information, contact your local dealer or GWInstek at www.gwinstek.com / [email protected].

Página 85 - UVP indicator

APPENDIX 173 APPENDIX Replacing the Dust Filter ... 175 Replace the Clock Battery ...

Página 86 - UnReg indicator

PEL-3000 User Manual 174 Other ...

Página 87

APPENDIX 175 Replacing the Dust Filter Background The dust filter should be replaced twice a year. Not replacing the filter will reduce performance

Página 88 - System Settings

PEL-3000 User Manual 176 Replace the Clock Battery Background The system clock keeps time using a user-replaceable battery. The battery should be

Página 89 - Display Settings

APPENDIX 177 GPIB Installation Background GPIB is an optional extra. The following instructions describe how to install the optional GPIB card if n

Página 90 - Language Settings

PEL-3000 User Manual 16 (PEL-3211 Booster Pack) LINK STBY Air Inlet The air inlet has a removable dust filter LCD display 3.5 inch LCD display Fu

Página 91 - Go-NoGo

PEL-3000 User Manual 178 PEL-3000 Default Settings The following default settings are the factory configuration settings for the PEL-3000. Main S

Página 92 - Running a Go-NoGo Test

APPENDIX 179 Main > Configure > Other Item Panel Settings Setup Memory Settings (all 100 sets) Soft Start OFF OFF Von Voltage 0.00V 0.00V Von

Página 93 - Test result

PEL-3000 User Manual 180 Main > Configure > Next Menu > Knob Item Panel Settings Setup Memory Settings (all 100 sets) Status Step Step CC

Página 94 - Program

APPENDIX 181 Frame Control Connector Contacts J1 Connector Pin name Pin number Description EXT R/V CONT 1 Used for voltage/resistance control of CC

Página 95

PEL-3000 User Manual 182 TRIG INPUT 11 When paused, clears the pause when a low level TTL signal is applied for 10 μs or longer. Pulled down the i

Página 96 - Create a Program

APPENDIX 183 J2 Connector Pin name Pin number Description N.C. 1 N.C. 2 N.C. 3 SUM I MON 4 Connect to SUM I MON of the J1 connector. PRL OUT+

Página 97

PEL-3000 User Manual 184 SUM I MON 4 Connected to SUM I MON of the J2 connector. PRL IN+ 5 Connected to PRL OUT+ of the J2 connector. PRL IN- 6 Co

Página 98

APPENDIX 185 J2 Connector Booster Pin name Pin number Description N.C. 1 N.C. 2 N.C. 3 SUM I MON 4 Connect to SUM I MON of the J1 connector. PRL

Página 99 - Create a Program Chain

PEL-3000 User Manual 186 Operating Mode Description CC Mode CC Mode When the unit is set to CC mode it will operate as a constant current load wh

Página 100 - Running a Program or Chain

APPENDIX 187 CC + CV ModeLoad InputVoltageLoad CurrentCV levelCC Current Value Note that when the source voltage is less than the CV level, no curre

Página 101 -  Press Exit[F5] to exit

GETTING STARTED 17 FileFUNCShift File (Shift + FUNC): Accesses the file system. Help/Utility Help Help: Access the help menu. UtilityHelpShift Uti

Página 102 - Sequence

PEL-3000 User Manual 188 CR+CV Mode When CR+CV mode is enabled, the unit will act as constant resistive load after the input voltage is greater th

Página 103

APPENDIX 189 CP ModeLoad CurrentLoad InputVoltageCP Value CP+CV Mode When CP+CV mode is enabled, the unit will act as a constant power load after th

Página 104 -  The number of available

PEL-3000 User Manual 190 CV Mode CV Mode When the unit is set to CV mode it will operate as a constant voltage load when connected to a constant

Página 105 - Start of step

APPENDIX 191 Operating Area PEL-3021

Página 106 - Timing Edit Configuration

PEL-3000 User Manual 192 PEL-3041

Página 107

APPENDIX 193 PEL-3111

Página 108 - Data Edit Configuration

PEL-3000 User Manual 194 PEL-3211 Booster Pack

Página 109

APPENDIX 195 PEL-3000 Specifications The specifications apply when the PEL-3000 is powered on for at least 30 minutes to warm-up to a temperature of

Página 110 - Running a Normal Sequence

PEL-3000 User Manual 196 CC Mode Model PEL-3021 PEL-3041 PEL-3111 Operating Range H Range 0A~35A 0A~70A 0A~210A M Range 0A~3.5A 0A~7A 0A~21A L Ra

Página 111 - Loop: 0001

APPENDIX 197 CR Mode Model PEL-3021 PEL-3041 PEL-3111 Operating Range*1 H Range 23.3336S~400uS (42.857mΩ~2.5kΩ) 46.6672S~800uS (21.428mΩ~1.25kΩ) 14

Página 112 - Fast Sequence Overview

This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied,

Página 113

PEL-3000 User Manual 18 Number pad P0P1P4P7CAL.P2P5P8LockP3P6P90147258369Clear Number pad: Used to enter numerical values. P0~P9 (Preset + Numbe

Página 114

PEL-3000 User Manual 198 Setting Range H Range 0V~157.5V M Range 0V~15.75V Resolution H Range 10mV M Range 1mV Accuracy of Setting*1 H, L Range ±(

Página 115

APPENDIX 199 Slew Rate Model PEL-3021 PEL-3041 PEL-3111 Setting Range (CC Mode) H Range 2.5mA/us~2.5A/us 5mA/us~5A/us 16mA/us~16A/us M Range 250uA/

Página 116

PEL-3000 User Manual 200 Ammeter H, M Range 0.000A~35.000A 0.000A~70.000A 0.00A~210.00A L Range 0.00mA~350.00mA 0.00mA~700.00mA 0.0mA~2100.0mA Acc

Página 117

APPENDIX 201 Slew Rate Setting Range (CR Mode) H Range 250uA/us~250mA/us 500uA/us~500mA/us 1.6mA/us~1.6A/us M Range 25uA/us~25mA/us 50uA/us~50mA/us

Página 118

PEL-3000 User Manual 202 Resistance Accuracy of setting H, M Range ±(0.5 % of set*1 + 0.5 % of f.s*2) + Vin*3/500 kΩ L Range ±(0.5 % of set*1 + 0.

Página 119 - Running a Fast Sequence

APPENDIX 203 Protection Function Model PEL-3021 PEL-3041 PEL-3111 Overvoltage protection(OVP) Turns off the load at 110% of the rated voltage Over

Página 120 - Run F.Seq

PEL-3000 User Manual 204 Fast Sequence Operation mode CC or CR Maximum number of steps 1000 Step Execution Time 25 μs – 600 ms Time resolution 1u

Página 121 - Save Recall

APPENDIX 205 Range Switch Input Switch ranges L, M, and H using a 2-bit signal Range Status Output Outputs range L, M, or H using 2-bit signal (op

Página 122 - File Types

PEL-3000 User Manual 206 Front Panel BNC Connector TRIG OUT Trigger output: Approx. 5V pulse width: Approx. 2 μs, output impedance: Approx. 500

Página 123

APPENDIX 207 PEL-3000 Dimensions PEL-3111 P0P1P4P7CAL.P2P5P8LockP3P6P9UtilityLocalFile0147258369EnterClearShiftPresetLoadOn/OffMainHelpFUNCShortTRI

Página 124 - Memory M256

GETTING STARTED 19 Front panel input terminals 1.5 - 150V175W0 - 35A Negative terminal. Positive terminal. IMON Out I MON OUT Current monitor BNC

Página 125 - Saving Files to USB Memory

PEL-3000 User Manual 208 PEL-3021, PEL-3041 I MON OUT TRIG OUT1.5 - 150V175W0 - 35AP0P1P4P7CAL.P2P5P8LockP3P6P9UtilityLocalFile0147258369Enter

Página 126

APPENDIX 209 PEL-3211 PEL-3211DC Electronic Load0 - 420A / 1.5 - 150V , 2100WLINK STBY427.75.05.8553.5592.5127.8147.8

Página 127

PEL-3000 User Manual 210 Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. No. 7-1, Jhongsing Rd, Tucheng Dist., New Taipei City 236, T

Página 128

INDEX 211 INDEX Accessories ... 12 Advanced configuration Alarm tone ... 87 Auto Loa

Página 129 - Path: usb:

PEL-3000 User Manual 212 Knob configuration Cursor ... 77 Step ...

Página 130 - Recall Memory Safety Setting

INDEX 213 About disassembly ... 4 Contact ... 172 Setting the date and time ... 2

Página 131 - File Utility

PEL-3000 User Manual 20 Rear Panel PEL-3021 / PEL-3041 47 - 63 Hz90 VA MAX.AC100 - 120 VAC200 - 240 VACFRAME CONTJ 1J 2SER. NO. LBRS232CGPIBWAR

Página 132 - Preset

GETTING STARTED 21 (PEL-3211 Booster Pack) RS232C Port GPIB USB B The USB B, RS232C and GPIB port are used for remote control. USB B port RS

Página 133 - Default Settings

PEL-3000 User Manual 22 Exhaust fan The exhaust fan is used to expel the heat from the unit. Please ensure there is at least 20cm distance between

Página 134 - User’s Default Setting

GETTING STARTED 23 Display ModeCVI RangeH 35AV RangeL 15VFunctionStaticConfigure0.000V0.000A0.00wSlewRate 2500.00CC B Value 0.000CC A Value 0.000LOA

Página 135 - EXTERNAL CONTROL

PEL-3000 User Manual 24 First Time Use Instructions Use the procedures below when first using the PEL-3000 to install the rack mount kit, power up

Página 136 - Analog Control

GETTING STARTED 25 GRA-413 (EIA standard) 128532.5482.6 GRA-413 (JIS standard) 149532.5482.6

Página 137 - Pin3 → EXT-V (-)

PEL-3000 User Manual 26 GRA-414-E (EIA standard) 132382.5477.6 GRA-414-J (JIS standard) 149382.5477.6

Página 138

GETTING STARTED 27 Power Up and Self Test Steps 1. Insert the AC power cord into the power socket. 2. Turn the external power switch on. (O →

Página 139

Table of Contents 1 Table of Contents SAFETY INSTRUCTIONS ... 3 GETTING STARTED ...

Página 140

PEL-3000 User Manual 28 Load Default Settings Description When first using the PEL-3000, recall the factory default settings to ensure the unit

Página 141 - Pin3 → EXT-R

GETTING STARTED 29 SystemInfoLoad Interface Time Set LOADRS23203/Sep/2012 Other9Hour 0Year 2012MinuteDayMonth93Date/TimeDate Load Wiring Wire Gau

Página 142

PEL-3000 User Manual 30 2 6.54304 0.51266 181 3 5.82676 0.64616 158 4 5.18922 0.81508 135 5 4.62026 1.02762 118 6 4.1148 1.29593 101 7 3.665

Página 143

GETTING STARTED 31 CurrentVoltagetimetime∆I∆TEMin. V The diagram above shows how changes in current can affect voltage. Limiting Load line induct

Página 144 -  See page 56 for CP mode

PEL-3000 User Manual 32 Load Wire Connections Description The PEL-3000 has input terminals on both the front and rear panels. Follow the procedur

Página 145

GETTING STARTED 33 Using the Front Panel Input Terminals Description The front panel input terminals feature polarity-distinct caps and accept M6 s

Página 146

PEL-3000 User Manual 34 Using the Rear Panel Input Terminals Description The rear panel input terminals accept up to M8-sized crimped terminals.

Página 147 - Load On/Off Status

GETTING STARTED 35 Using the Terminal Cover Description The rear panel terminal cover should be used to prevent electric shock. The rear panel term

Página 148 - I Range Status

PEL-3000 User Manual 36 2Detail3 4. Use your thumb to slide the terminal covers shut, as shown in the diagram below. 4

Página 149 - External Trigger Signal

GETTING STARTED 37 5. When the top and bottom covers are flush, reinsert the screw that was removed in step 1. 5

Página 150

PEL-3000 User Manual 2 Replace the Clock Battery ... 176 GPIB Installation ... 177 PEL-3000 Default S

Página 151 - Short Control

PEL-3000 User Manual 38 Remote Sense Description Remote sense can be used to help compensate for long cable length. The longer the cable, the hig

Página 152 - Monitor Signal Output

GETTING STARTED 39 Firmware Update Description The PEL-3000 allows the firmware to be updated by end-users. Before using the PEL-3000, please check

Página 153 - Monitor Output Range

PEL-3000 User Manual 40 SystemInfoLoad Interface Time Set LOADRS23203/Sep/2012Model : PEL-30XX Firmware Ver. : 1.01.011Serial Number : 00000001O

Página 154 - Parallel Operation

GETTING STARTED 41 Conventions The following conventions are used throughout the user manual. Read the conventions below for a basic grasp of how to

Página 155 - Connection

PEL-3000 User Manual 42 Toggle Parameter or State ModeCCFunction/ItemParameter or State This type of soft-menu icon has the function/item on the t

Página 156 - No ferrite core to

GETTING STARTED 43 Parameter Input The scroll wheel, Enter key and number pad can be used to edit parameter values. Number padScroll wheelEnter key

Página 157 - Configuration

PEL-3000 User Manual 44 3. Then use the number pad* or scroll wheel** to edit the parameter value. ModeCCI RangeH 35AV RangeL 15VFunctionStati

Página 158

GETTING STARTED 45 Step Mode: This is the default step resolution method and will only be available to use when it is applicable (Indicated by Fine

Página 159 - Turning the Load On

PEL-3000 User Manual 46 Entering Alphanumeric Characters When renaming files, creating memos or notes, you will be required to enter alphanumeric

Página 160 - Disable Parallel Mode

GETTING STARTED 47 Help Menu When any function key has been pressed or when a menu has been opened, the HELP key can be used to display a detailed d

Página 161 - REMOTE CONTROL

SAFETY INSTRUCTIONS 3 SAFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow during operation and storage. R

Página 162 - Interface Configuration

PEL-3000 User Manual 48 OPERATION Basic Operation ... 51 CC Mode ...

Página 163

OPERATION 49 Cursor Mode Configuration ... 77 Step Mode Configuratio

Página 164 - 1 2 3 4 5

PEL-3000 User Manual 50 File Types ... 120 Saving F

Página 165

OPERATION 51 Basic Operation The PEL-3000 supports 7 main operating modes: CC, CC+CV; CR, CR+CV; CV; CP, CP+CV CC Mode Description In Constant Cu

Página 166

PEL-3000 User Manual 52 6. Set the current level parameters using the scroll wheel and number pad.  For Static mode, set CC A Value and/or CC

Página 167

OPERATION 53 CR Mode Description In Constant Resistance Mode, the unit will maintain a constant resistive load by varying the current. CR mode use

Página 168

PEL-3000 User Manual 54 6. Set the resistance or conductance level parameters using the scroll wheel and number pad.  For Static mode, set CR

Página 169

OPERATION 55 CR Units Description The CR setting units can be set to ohm (Ω) or millisiemens (mS). Operation 1. Make sure the load is off. 2. P

Página 170 - GPIB Function Check

PEL-3000 User Manual 56 6. Set the voltage level parameters using the scroll wheel and number pad.  Set CV A Value and/or CV B Value.  The

Página 171

OPERATION 57 Warning If you change the mode or the range when the load is already on, the load will be turned off automatically. Operation 1. Mak

Página 172

PEL-3000 User Manual 4 Safety Guidelines General Guideline CAUTION  Do not place any heavy object on the instrument. Note: Only 2 units can be

Página 173 - The load won’t turn on

PEL-3000 User Manual 58 Display ModeCPI RangeH 35AV RangeH 150VFunctionStaticConfigure0.00V0.000A0.00wCP B Value 0.00CP A Value 0.00LOADRS232A Val

Página 174 - PEL-3000 User Manual

OPERATION 59 Display ModeCPI RangeH 35AV RangeL 15VFunctionDynamicConfigure0.000V0.000A0.00w+CV 5.500Timer2 0.025Timer1 0.025LOADRS23203/Sep/2012Vms

Página 175 - APPENDIX

PEL-3000 User Manual 60 Note  The load can be set to automatically turn on at start up. See page 75.  The load can be turned on via remote co

Página 176

OPERATION 61 Operation 1. The short function can be turned on and off by pressing the Short key.  The Short key will turn red when the short func

Página 177 - Replacing the Dust Filter

PEL-3000 User Manual 62 Operation 1. The keys can be locked and unlocked by pressing Shift+LockClear.  LOCK will appear in the Mainframe status

Página 178 - Replace the Clock Battery

OPERATION 63 Basic Configuration The basic configuration settings are the common configuration settings that are used for each operating mode. After

Página 179 - - potential

PEL-3000 User Manual 64 LevelTimeLevel 2Level 1Timer1 Timer2 Timer1Dynamic Mode Note Dynamic mode is not available for CV mode. Operation 1. Mak

Página 180 - PEL-3000 Default Settings

OPERATION 65  The load can be ―on‖ when switching between A Value and B Value. Display: Static Mode ModeCPI RangeH 35AV RangeH 150VFunctionSt

Página 181

PEL-3000 User Manual 66 Operation 1. Make sure the load is off. 2. Press Main > Configure[F5] > Other[F2] and set the Dyna. Level setting

Página 182

OPERATION 67 Select the Switching Time Configuration for Dynamic Mode Description The switching time for dynamic mode can be configured to switch b

Página 183

SAFETY INSTRUCTIONS 5 Power Supply WARNING  AC Input voltage range: 100~120VAC/200~240VAC (90~132VAC/180~250VAC)  Frequency: 47~63Hz  Power:

Página 184

PEL-3000 User Manual 68 Display ModeCCI RangeH 35AV RangeH 150VFunctionDynamicConfigure0.00V0.000A0.00wTimer1 0.025SlewRate 2500.00SlewRate 2500.0

Página 185

OPERATION 69 Display ModeCVI RangeH 35AV RangeL 15VResponseSlowConfigure0.000V0.000A0.00wCV B Value 15.000CV A Value 15.000LOADRS232A ValueFine03/Se

Página 186

PEL-3000 User Manual 70 Advanced Configuration Settings Use the advanced configuration settings to configure settings other than those described i

Página 187

OPERATION 71 Von Voltage Settings Von Voltage Level Description The Von Voltage is the threshold voltage at which the load module will start to sin

Página 188 - Operating Mode Description

PEL-3000 User Manual 72 Von Voltage Latch Description When Von Latch is set to ON, the load will continue to sink current after being ―latched‖,

Página 189 - CR Mode

OPERATION 73 Von Voltage Delay Description Von Delay is the amount of time the unit will wait before turning the load on after the Von Voltage thre

Página 190 - CP Mode

PEL-3000 User Manual 74 Operation 1. Press Main > Configure[F5] > Other[F2] and turn the Count Time on or off. Range: ON, OFF Display Mo

Página 191

OPERATION 75 Auto Load Configuration Description The PEL-3000 can be configured to automatically load the last program, normal sequence, fast seque

Página 192 - CV level

PEL-3000 User Manual 76 Load Off (Mode) and Load Off (Range) Description By default the load will automatically turn off when the either the oper

Página 193 - Operating Area

OPERATION 77 Step Resolution Configuration There are two different ways to set the set resolution when using the scroll wheel to edit parameters. St

Página 194

PEL-3000 User Manual 6 (Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The instrument falls

Página 195

PEL-3000 User Manual 78 Step Mode Configuration Description When set to Step Mode, the voltage, current, resistance and power settings can have t

Página 196 - Booster Pack

OPERATION 79 Operation 1. Press Main > Configure[F5] > Next Menu[F4] > Knob[F2] and make sure the Status setting is set to Step. 2. Set t

Página 197 - PEL-3000 Specifications

PEL-3000 User Manual 80 Protection Settings The Protection settings are used to prevent damage to the unit or the DUT by excessive current, voltag

Página 198 - CC Mode

OPERATION 81 Display 2.596V500.06mA1.2982wCV B Value 15.000CV A Value 15.000LOADRS232A ValueFine03/Sep/2012VVOCP AlarmPlease Press EnterTo Clear Ala

Página 199 - CV Mode

PEL-3000 User Manual 82 Display 2.596V500.06mA1.2982wCV B Value 15.000CV A Value 15.000LOADRS232A ValueFine03/Sep/2012VVOPP AlarmPlease Press Ente

Página 200

OPERATION 83 Display 2.596V500.06mA1.2982wCV B Value 15.000CV A Value 15.000LOADRS232A ValueFine03/Sep/2012VVUVP AlarmPlease Press EnterTo Clear Ala

Página 201 - Slew Rate

PEL-3000 User Manual 84 Display 2.596V500.06mA1.2982wCV B Value 15.000CV A Value 15.000LOADRS232A ValueFine03/Sep/2012VVOVP AlarmPlease Press Ente

Página 202 - Dynamic Mode

OPERATION 85 Para Description The Para error message will appear on the display when the PEL-3000 is used in parallel and if an error is produced.

Página 203 - APPENDIX

PEL-3000 User Manual 86 System Settings The following section covers a number or miscellaneous system settings such as:  Speaker settings  Dis

Página 204 - Remote Sensing

OPERATION 87 Alarm Tone Settings Description The alarm tone for the unit can be turned on or off in the utility menu. The alarm tone can be set sep

Página 205 - Protection Function

SAFETY INSTRUCTIONS 7 Power cord for the United Kingdom When using the instrument in the United Kingdom, make sure the power cord meets the followin

Página 206 - Analog External Control

PEL-3000 User Manual 88 Control Settings Description The Knob Type setting determines if values are updated immediately as they are edited or if

Página 207

OPERATION 89 Go-NoGo The Go-NoGo configuration is used to create pass/fail limits on the voltage or current input. If the voltage/current exceeds t

Página 208 - Front Panel BNC Connector

PEL-3000 User Manual 90 5. Set the Delay Time.  The delay time setting will delay activating the Go-NoGo testing by a specified amount of time

Página 209 - PEL-3000 Dimensions

OPERATION 91 Display: GO ModeCPIRangeH 35AVRangeL 15VFunctionDynamicConfigure0.000V0.000A0.00wTimer1 0.025Level2 0.00Level1 0.00LOADRS232Fine03/Sep/

Página 210 - PEL-3021, PEL-3041

PEL-3000 User Manual 92 Program The PEL-3000 can create programs that are designed to step-through up to 16 pre-set load operations. The program

Página 211

OPERATION 93 program chains.  Program chains need not be executed in order.  There are 16 steps to a program.  There are up to 16 programs to

Página 212 - Declaration of Conformity

PEL-3000 User Manual 94 Timing Diagram for Single Step Below is a timing diagram of a single step in a program. On-Time Off-TimeStep test timeStar

Página 213 - INDEX

OPERATION 95 PROG 01 ~ 16 3. Select a STEP in the selected program. STEP 01 ~ 16 4. Select Memory and select which memory location to load fo

Página 214

PEL-3000 User Manual 96 8. Choose the P/F-Time (pass/fail time) in seconds.  The P/F-Time refers to the P/F delay time. This delay time includ

Página 215

OPERATION 97 Create a Program Chain Note Before creating a program chain, make sure a number of programs have already been saved. These will be us