REMOTE CONTROL
75
Trigger Commands
The trigger commands generate and configure software triggers.
TRIGger:TRANsient[:IMMediate] ...................................... 75
TRIGger:TRANsient:SOURce ............................................ 75
TRIGger:OUTPut[:IMMediate] .......................................... 76
TRIGger:OUTPut:SOURce ................................................. 76
Trigger Command Examples ............................................... 76
TRIGger:TRANsient[:IMMediate]
Generates a software trigger for the transient
trigger system. On a trigger, sets the voltage &
current. Refer to the :CURR:TRIG and VOLT:TRIG
commands on page 69 and 73, respectively.
TRIGger:TRANsient[:IMMediate]
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
Sets or queries the trigger source for the transient
system.
TRIGger:TRANsient:SOURce {BUS | IMMediate}
TRIGger:TRANsient:SOURce?
Internal software trigger. Waits for the
*TRG (or IEEE 488.1 “get” group
execute trigger) command to start the
trigger.
Starts the trigger immediately.
(default)
TRIG:TRAN:SOUR BUS