COMMAND DETAILS
137
Return Parameter
Returns the address type
Example
:TRIGger:BUS:B1:I2C:ADDRess:TYPe?
CBUS
:TRIGger:BUS:B1:I2C:ADDRess:VALue
Description
Sets or queries the I
2
C bus address value when the
I
2
C bus is set to trigger on Address or
Address/Data.
Syntax
:TRIGger:BUS:B1:I2C:ADDRess:VALue {string | ? }
Related
commands
:TRIGger:BUS:B1:I2C:ADDRess:MODe
Parameter
<sting>
7/10 characters, must be enclosed in
double quotes, "string".
x = don’t care
1 = binary 1
0 = binary 0
Return Parameter
Returns the address value.
Example1
:TRIGger:BUS:B1:I2C:ADDRess:VALue "xxx0101"
Sets the address to XXX0101
Example 2
:TRIGger:BUS:B1:I2C:ADDRess:VALue?
XXX0101
:TRIGger:BUS:B1:I2C:ADDRess:DIRection
Description
Sets or queries the address bit as read write or
don’t care.
Note
This setting only applies when the I
2
C trigger is set
to trigger on Address or Address/Data
Syntax
:TRIGger:BUS:B1:I2C:ADDRess:DIRection { READ |
WRITE | NOCARE | ? }
Set
Query
Set
Query