Digital Storage Oscilloscope
GDS-3000 Series
PROGRAMMING MANUAL
GW INSTEK PART NO. 82DS-33040IC1
ISO-9001 CERTIFIED MANUFACTURER
Vista de página 0
1 2 3 4 5 6 ... 102 103

Resumo do Conteúdo

Página 1 - Digital Storage Oscilloscope

Digital Storage Oscilloscope GDS-3000 Series PROGRAMMING MANUAL GW INSTEK PART NO. 82DS-33040IC1 ISO-9001 CERTIFIED MANUFACTURER

Página 2

GDS-3000 Programming Manual 10 Configure GPIB Interface To use GPIB, the optional GPIB to USB (GUG-001) adapter must be used. The GPIB address ca

Página 3 - Table of Contents

GDS-3000 Programming Manual 100 :GONogo:SCRipt Set Description Activates/Deactivates the Go/NoGo APP or queries its state. Syntax :GONogo:SCRip

Página 4 - INTERFACE OVERVIEW

COMMAND DETAILS 101 Example :TEMPlate:MAXimum REF1 Saves the maximum template to REF1. :TEMPlate:MINimum SetQuery Description Defines or queries

Página 5 - Interface Configuration

GDS-3000 Programming Manual 102 Parameter <NR2> Desired template position (-12.0 ~ +12.0 divisions) Return parameter Returns the position in

Página 7 - 1 2 3 4 5

INTERFACE OVERVIEW 11 USB/RS-232C Remote Control Software Terminal application (USB/RS-232C) Invoke the terminal application such as hyper termin

Página 8

GDS-3000 Programming Manual 12 Using the Ethernet Interface Background The following example will show you how to use the Ethernet interface to a

Página 9 - Left Right

INTERFACE OVERVIEW 13 4. The Local Area Connection Status panel appears. Click the Properties icon.

Página 10 - Configure GPIB Interface

GDS-3000 Programming Manual 14 5. The Networking tab will be shown. 6. Double click on the Internet Protocol Version 4 (TCP/IPv4) option in the

Página 11

INTERFACE OVERVIEW 15 7. The Internet Protocol Version 4 (TCP/IPv4) Properties will appear. Manually set the IP address. (Click the Use the fo

Página 12 - Using the Ethernet Interface

GDS-3000 Programming Manual 16 11. Set the DSO Ethernet settings. See page 7 for details. 12. Press the Utility key. Utility 13. Press I/O fr

Página 13

INTERFACE OVERVIEW 17 18. To check to see if the Ethernet settings are working enter the DSO’s IP address into a browser: 172.16.1.3 19. Nex

Página 14

GDS-3000 Programming Manual 18 20. Right-click Devices and Interfaces and select Create New.

Página 15

INTERFACE OVERVIEW 19 21. Select VISA TCP/IP Resource and click Next. 22. Select Manual Entry or Raw Socket.

Página 16

This manual contains proprietary information which is protected by copyright. All rights are reserved. No part of this manual may be photocopied,

Página 17

GDS-3000 Programming Manual 20 23. Enter the IP address and Socket number of the DSO. Press Validate. 24. When Validate has been clicked, the

Página 18

INTERFACE OVERVIEW 21 26. The DSO IP address and socket will now appear under Network Devices in the My System list. 27. Click on the Open Visa T

Página 19

GDS-3000 Programming Manual 22 29. Check Enable Termination Character and set the Termination Character to Line Feed - \n with a value of 0x0A 30

Página 20

COMMAND OVERVIEW 23 COMMAND OVERVIEW The Command overview chapter lists all GDS-3000 commands in functional order as well as alphabetical order. T

Página 21

GDS-3000 Programming Manual 24 Command format :TIMebase:SCALe <NR3>LF1 2 3 4 1: command header 2: single space 3: parameter 4: message termi

Página 22

COMMAND OVERVIEW 25 List of Commands in Functional Order Common *IDN? ...

Página 23 - COMMAND OVERVIEW

GDS-3000 Programming Manual 26 Math :MATH:DISP ... 42 :MATH:TYPe ...

Página 24 - :TIMebase:SCALe <NR3>LF

COMMAND OVERVIEW 27 Display :DISPlay:INTensity:WAVEform ... 54 :DISPlay:INTensity:GRATicule...

Página 25

GDS-3000 Programming Manual 28 :MEASure:FFRDelay ... 68 :MEASure:FFFDelay ...

Página 26

COMMAND OVERVIEW 29 :TRIGger:NREJ ... 82 :TRIGger:REJect ...

Página 27

TABLE OF CONTENTS 3 Table of Contents INTERFACE OVERVIEW ... 4 Front Panel Overview ...

Página 28

GDS-3000 Programming Manual 30 :GONogo:EXECute ... 98 :GONogo:FUNCtion ...

Página 29

COMMAND DETAILS 31 COMMAND DETAILS The Command details chapter shows the detailed syntax, equivalent panel operation, and example for each command

Página 30

GDS-3000 Programming Manual 32 Common Commands *IDN? ... 32 *LRN?...

Página 31 - COMMAND DETAILS

COMMAND DETAILS 33 *RCL Set Description Recalls a set of panel setting. Syntax *RCL {1 | 2 | 3 |…. | 20} Example *RCL 1 Recalls setup from set 1

Página 32 - Common Commands

GDS-3000 Programming Manual 34 Syntax :ACQuire:AVERage {2 | 4 | 8 | 16 | 32 | 64 | 128 | 256 | ?} Parameter 2, 4, 8 ,16, 32, 64, 128, 256 Note Bef

Página 33 - Acquisition Commands

COMMAND DETAILS 35 Example :ACQuire1:MEMory? Memory Length,25000;IntpDistance,0;Trigger Address,12499;Trigger Level,1.00V;Sou rce,CH1;Vertical Uni

Página 34 - :ACQuire<x>:MEMory?

GDS-3000 Programming Manual 36 Autoscale Commands :AUTOSet ... 36 :AUTORange .

Página 35 - :ACQuire<x>:STATe?

COMMAND DETAILS 37 :CHANnel<x>:BWLimit SetQuery Description Selects or returns the bandwidth limit on/off. Syntax :CHANnel<x>:BWLimit

Página 36 - Vertical Commands

GDS-3000 Programming Manual 38 Syntax :CHANnel<x>:DESKew { <NR3> | ?} Parameter <x> Channel <NR3> Deskew time 1/2/3/4 CH1

Página 37

COMMAND DETAILS 39 :CHANnel<x>:IMPedance SetQuery Description Sets the impedance of the oscilloscope. Syntax :CHANnel<x>:IMPedance {

Página 38

GDS-3000 Programming Manual 4 INTERFACE OVERVIEW This manual describes how to use the GDS-3000’s remote command functionality and lists the comman

Página 39

GDS-3000 Programming Manual 40 Parameter <x> Channel <NRf> Position 1/2/3/4 CH1/2 /3/4 Range depends on the vertical scale. Return p

Página 40

COMMAND DETAILS 41 Return parameter Returns the probe type. Example :CHANnel1:PROBe:TYPe VOLTage Sets the Channel 1 probe type to voltage. :CHANne

Página 41 - Math Commands

GDS-3000 Programming Manual 42 :MATH:DISP SetQuery Description Turns the math display on or off on the screen. Note: Math display cannot be used

Página 42

COMMAND DETAILS 43 Parameter <x> Source number: 1/2/3/4 CH1~4 Channel 1 to 4 REF1~4 Reference waveforms 1 to 4 Return parameter Returns th

Página 43

GDS-3000 Programming Manual 44 Example :MATH:DUAL:POSition 1.0E+0 Sets the vertical position to 1.00 unit/div. :Math:DUAL:POSition? 1.0E+0 Returns

Página 44

COMMAND DETAILS 45 Example :MATH:FFT:SOURce CH1 Sets the FFT math source as channel 1. :MATH:FFT:SOURce? CH1 Return the FFT math source as channe

Página 45

GDS-3000 Programming Manual 46 :MATH:FFT:POSition SetQuery Description Sets the vertical position of the displayed FFT result. Syntax MATH:FFT:

Página 46

COMMAND DETAILS 47 Cursor Commands :CURSor:MODe ... 47 :CURSor:SOURce ...

Página 47 - Cursor Commands

GDS-3000 Programming Manual 48 Example :CURSor:MODe OFF Turns the cursors off. :CURSor:SOURce SetQuery Description Sets or queries the cursor so

Página 48

COMMAND DETAILS 49 Example :CURSor:H2Position 1.5E-3 Sets the H2 cursor position to 1.5ms. :CURSor:HDELta Query Description Returns the delta of

Página 49

INTERFACE OVERVIEW 5 Interface Configuration Configure USB Interface USB configuration PC side connector Type A, host GDS-3000 side connector Typ

Página 50

GDS-3000 Programming Manual 50 Example :CURSor:V2Position 1.1E-1 Sets the V2 cursor position to 110mA. :CURSor:VDELta Query Description Returns

Página 51

COMMAND DETAILS 51 Return Parameter <NR3> Returns the delta value of cursor 1 and 2 as <NR3>. Example :CURSor:XY:RECTangular:X:DELta?

Página 52

GDS-3000 Programming Manual 52 :CURSor:XY:POLar:THETA:POSition<x> Query Description Queries the theta in XY mode for the specified cursor,

Página 53

COMMAND DETAILS 53 Return parameter Returns the product value of the Cursor1 or Cursor2 as <NR3>. Example :CURSor:XY:PRODuct:POSition1? 9.44

Página 54 - Display Commands

GDS-3000 Programming Manual 54 Return parameter Returns the ratio delta as <NR3> V/A| V/V|A/A Example :CURSor:XY:RATio:DELta? 5.39E+1 Return

Página 55

COMMAND DETAILS 55 Example :DISPlay:INTensity:GRATicule 5.0E+1 Sets the graticule intensity to 50%. :DISPlay:PERSistence SetQuery Description Se

Página 56 - Hardcopy Commands

GDS-3000 Programming Manual 56 :DISPlay:WAVEform SetQuery Description Sets or queries whether the waveforms are drawn as vectors or dots. Syn

Página 57

COMMAND DETAILS 57 Example :HARDcopy:MODe PRINT Sets hardcopy to print. :HARDcopy:PRINTINKSaver SetQuery Description Sets Inksaver On or Off for

Página 58 - Measure Commands

GDS-3000 Programming Manual 58 Measure Commands :MEASure:GATing ... 59 :MEASure:SOURce<x

Página 59

COMMAND DETAILS 59 :MEASure:GATing SetQuery Description Sets or queries the measurement gating. Syntax :MEASure:GATing {OFF|SCREen|CURSor|?} P

Página 60 - :MEASure:FPReshoot

GDS-3000 Programming Manual 6 Configure RS-232C Interface RS-232C configuration Connector DB-9, Male Baud rate 2400, 4800, 9600, 19200, 38400, 57

Página 61

GDS-3000 Programming Manual 60 Note Before using this command, select the measurement channel. See the example below. Example :MEASure:SOURce CH1

Página 62 - :MEASure:PWIDth

COMMAND DETAILS 61 :MEASure:FREQuency Query Description Returns the frequency value. Syntax :MEASure:FREQuency{?} Return parameter Returns the f

Página 63

GDS-3000 Programming Manual 62 Note Before using this command, select the measurement channel. See the example below. Example :MEASure:SOURce1 CH

Página 64 - :MEASure:AVERage

COMMAND DETAILS 63 :MEASure:RISe Query Description Returns the first pulse rise time. Syntax :MEASure:RISe{?} Return parameter Returns the rise

Página 65 - :MEASure:LOW

GDS-3000 Programming Manual 64 Return parameter Returns the riser overshoot as <NR3> Note Before using this command, select the measurement

Página 66

COMMAND DETAILS 65 Example :MEASure:SOURce1 CH 1 :MEASure:AVERage? 1.82E-3 Selects Channel 1, and then measures the average value. :MEASure:HIGH Q

Página 67 - :MEASure:FRRDelay

GDS-3000 Programming Manual 66 :MEASure: MAX Query Description Returns the maximum amplitude. Syntax :MEASure:MAX{?} Return parameter Returns th

Página 68 - :MEASure:FFRDelay

COMMAND DETAILS 67 Return parameter Returns the voltage or current peak to peak measurement as <NR3>. Note Before using this command, select

Página 69 - :MEASure:LRRDelay

GDS-3000 Programming Manual 68 Example :MEASure:SOURce1 CH 1 :MEASure:SOURce2 CH 2 :MEASure:FRRDelay? -4.68E-6 Select channel 1 and 2 as source1/2

Página 70 - :MEASure:LFRDelay

COMMAND DETAILS 69 Example :MEASure:SOURce1 CH 1 :MEASure:SOURce2 CH 2 :MEASure:FRRDelay? -8.56E-6 Select channel 1 and 2 as delay source1/2, and

Página 71 - :MEASure:PHAse

INTERFACE OVERVIEW 7 7. Press Save Now to save the settings. 8. Connect the RS-232C cable to the rear panel port: DB-9 male connector. For a

Página 72 - Reference Commands

GDS-3000 Programming Manual 70 Example :MEASure:SOURce1 CH 1 :MEASure:SOURce2 CH 2 : MEASure:LRRDelay? -8.89E-6 Select channel 1 and 2 as delay so

Página 73

COMMAND DETAILS 71 Example :MEASure:SOURce1 CH 1 :MEASure:SOURce2 CH 2 :MEASure:LFRDelay? -9.99E-6 Select channel 1 and 2 as delay source1/2, and

Página 74 - Return parameter

GDS-3000 Programming Manual 72 Example :MEASure:SOURce1 CH 1 :MEASure:SOURce2 CH 2 :MEASure:PHAse? 4.50E+01 Select channel 1 and 2 as phase source

Página 75 - Single Command

COMMAND DETAILS 73 :REF<x>:TIMebase:POSition SetQuery Description Sets or returns the selected reference waveform time base position. Synt

Página 76 - Split Window Command

GDS-3000 Programming Manual 74 :REF<x>:OFFSet SetQuery Description Sets or returns the selected reference waveform vertical position (offs

Página 77 - Timebase Commands

COMMAND DETAILS 75 Run Command :RUN Set Description The run command allows the oscilloscope to continuously make acquisitions (equivalent to pre

Página 78 - :TIMebase:WINDow:POSition

GDS-3000 Programming Manual 76 Force Command :FORCe Set Description The Force command forces an acquisition. (equivalent to pressing the Force k

Página 79 - :TIMebase:WINDow:SCALe

COMMAND DETAILS 77 Timebase Commands :TIMebase:POSition ... 77 :TIMebase:SCALe ...

Página 80 - Trigger Commands

GDS-3000 Programming Manual 78 Syntax :TIMebase:MODe {MAIN | WINDow | SPLIT | XY | ?} Parameter MAIN Sets the time base mode to main screen. WIND

Página 81

COMMAND DETAILS 79 :TIMebase:WINDow:SCALe SetQuery Description Sets or queries the zoom horizontal scale. Syntax :TIMebase:WINDow:SCALe {<N

Página 82

GDS-3000 Programming Manual 8 Background The Ethernet interface is used for remote control over a network and to retrieve the scope’s screen shot,

Página 83

GDS-3000 Programming Manual 80 Trigger Commands :TRIGger:FREQuency ... 81 :TRIGger:TYPe ...

Página 84

COMMAND DETAILS 81 :TRIGger:FREQuency Query Description Queries the trigger frequency. Syntax :TRIGger:FREQuency{?} Return parameter Returns t

Página 85

GDS-3000 Programming Manual 82 EXT External source LINe AC Line Return parameter Returns the trigger source. Example :TRIGger:SOURce CH1 Sets t

Página 86

COMMAND DETAILS 83 Parameter OFF Frequency rejection off. HF High frequency filter on LF Low frequency filter on Return parameter Returns the st

Página 87

GDS-3000 Programming Manual 84 Syntax :TRIGger:HLEVel {<NRf> | ?} Parameter <NRf> High level value Return parameter Returns the trigge

Página 88

COMMAND DETAILS 85 Parameter TIMe Sets the delay type to time. EVENt Sets the delay type to event. Return parameter Returns the trigger delay typ

Página 89

GDS-3000 Programming Manual 86 Syntax :TRIGger:DELay:LEVel {<NRf> | ?} Parameter <NRf> Delay trigger level Return parameter Returns th

Página 90

COMMAND DETAILS 87 :TRIGger:RISEFall :SLOP SetQuery Description Sets or queries the Rise & Fall slope. Syntax :TRIGger:RISEFall :SLOP {RISe

Página 91 - System Commands

GDS-3000 Programming Manual 88 :TRIGger:VIDeo:FIELd SetQuery Description Sets or queries the video trigger field. Syntax :TRIGger:VIDeo:FIELd {F

Página 92 - Save/Recall Commands

COMMAND DETAILS 89 Example :TRIGger:VIDeo:POLarity POSitive Sets the video trigger polarity to positive. :TRIGger:PULSe:WHEn SetQuery Descripti

Página 93 - GDS-3000 series

INTERFACE OVERVIEW 9 5. Use the Up and Down arrows on the side menu to navigate each Ethernet configuration item. Items MAC Address, Instrum

Página 94

GDS-3000 Programming Manual 90 :TRIGger:ALTernate SetQuery Description Sets alternating between source triggers on or off. Syntax :TRIGger:ALTer

Página 95 - :SAVe:WAVEform

COMMAND DETAILS 91 EITher Sets the triggering point as either stays high or stays low. Example :TRIGger:TIMEOut:WHEn HIGH Sets the triggering po

Página 96 - SPREADSheet

GDS-3000 Programming Manual 92 Example :SYSTem:LOCK ON Turns the panel lock on. Save/Recall Commands :RECAll:SETUp ...

Página 97 - Go_NoGo Commands

COMMAND DETAILS 93 :RECAll:WAVEform W<n>,REF<x> SetQuery Description Recalls a waveform from wave1~wave20 or from file to REF1~4. Sy

Página 98

GDS-3000 Programming Manual 94 :SAVe:IMAGe:FILEFormat SetQuery Description Sets the file format for image. Syntax :SAVe:IMAGe:FILEFormat {PNG |

Página 99

COMMAND DETAILS 95 Example :SAVe:SETUp S1 Saves the current setup to Set1 in internal memory. :SAVe:SETUp “Disk:/DS0001.SET” Saves the current set

Página 100

GDS-3000 Programming Manual 96 Example :SAVe:WAVEform CH1, REF2 Saves the channel1 waveform to REF2. :SAVe:WAVEform ALL, “Disk:/ALL001” Creates a

Página 101

COMMAND DETAILS 97 Go_NoGo Commands Note: The Go_NoGo and Template commands can only be used with firmware version V1.19 or later. In addition, th

Página 102

GDS-3000 Programming Manual 98 :GONogo:EXECute SetQuery Description Enables or disables the Go/NoGo function or queries its state. Syntax :GONogo

Página 103 - COMMAND DETAILS

COMMAND DETAILS 99 Parameter/ Return Parameter EXITs Sets the NoGo condition to when the input signal exceeds the limit boundary. ENTers Sets the