COMMAND DETAILS
247
:SEGMents:SAVe:SELect:STARt
Description
Sets or queries the starting segment to save from.
Syntax
:SEGMents:SAVe:SELect:STARt
{SETTOMIN | SETTOMAX | <NR1> | ? }
Related
commands
:SEGMents:TOTalnum
Parameter/
Return parameter
SETTOMIN
Sets the starting segment to min
segment.
SETTOMAX
Sets the starting segment to the max
segment.
<NR1>
Sets the segment to 1~2048.
Example
:SEGMents:SAVe:SELect:STARt 2
Sets the starting segment to segment number 2.
:SEGMents:SAVe:SELect:END
Description
Sets or queries the ending segment to save from.
Syntax
:SEGMents:SAVe:SELect:END
{SETTOMIN | SETTOMAX | <NR1> | ? }
Related
commands
:SEGMents:TOTalnum
Parameter/
Return parameter
SETTOMIN
Sets the starting segment to min
segment.
SETTOMAX
Sets the starting segment to the max
segment.
<NR1>
Sets the segment to 1~2048.
Example
:SEGMents:SAVe:SELect:END 10
Sets the ending segment to segment number 10.