Sets alternating between source triggers on or off.
:TRIGger:ALTernate {OFF | ON |?}
Returns the Alternate trigger status (ON, OFF).
:TRIGger:ALTernate ON
Turns on alternating between source triggers.
Sets or queries the trigger level.
:TRIGger:LEVel {<NRf> | ?}
Returns the trigger level as <NR3>.
:TRIGger:LEVel 3.30E-3
Sets the trigger level to 330mV/mA.
Note: This command is equal to :TRIGger:HLEVel
Sets or queries the timeout trigger triggering point.
Note: Only applicable for 500MHz models with the
Timeout trigger APP installed.
:TRIGger:TIMEOut:WHEn {HIGH | LOW | EITher | ?}
Parameter/
Return parameter
Sets the triggering point as stays high.
Sets the triggering point as stays low.