COMMAND DETAILS
127
Query Example
:STAT:QUES:COND?
2
OV (overvoltage) error.
:STATus:QUEStionable:ENABle
Channel Specific
Command
Description
Sets which events are enabled in the Questionable
Status Enable register. The mask values are the bit
weights of the events. See page 166 for details.
Syntax
:STATus:QUEStionable:ENABle <NR1>
Parameter
<NR1>
Event
<NR1>
Event
1
OC
16
OT
2
OV
32
G/N
4
OP
64
UVP
8
RV
128~65535
Not used
Example
:STAT:QUES:ENAB 12
Events OP (Bit 3) and RV
(Bit 4) are enabled in the
Questionable Status
Enable register.
Query Syntax
:STATus:QUEStionable:ENABle? <NR1>
Return Parameter
<NR1>
Event
<NR1>
Event
1
OC
16
OT
2
OV
32
G/N
4
OP
64
UVP
8
RV
128~65535
Not used
Query Example
:STAT:QUES:ENAB?
4
The OP event is enabled.
:STATus:QUEStionable[:EVENt]?
Channel Specific
Command
Description
Returns the status of the Questionable Status Event
register. The Questionable Status Event register is
cleared upon reading.
Vista de página 126