Gw-instek PEL-2000 Series Programming User Manual Manual do Utilizador Página 151

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 170
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 150
COMMAND DETAILS
151
1 unit = 1 ohm
Returns the resistance
value.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:SEQ:POIN:RES?
100
The resistance value is
100 ohm.
:SEQuence:POINt:CURRent
Channel Specific
Command
Description
Sets the current value for the current point. CC
mode only.
Syntax
:SEQuence:POINt:CURRent <NRf>[A]|MIN|MAX
Parameter
<NRf>[A], MIN, MAX
Current value
10
10A
100 A
100A
MAX/MIN
Maximum or minimum
value.
Example
:SEQ:POIN:CURR 1
Sets the current to 1.
Query Syntax
:SEQuence:POINt::CURRent? [MAX|MIN]
Return Parameter
<NR1>
Current Value
1 unit = 1 A
Returns the current value.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:SEQ:POIN:CURR?
1
The current value is 1
amp.
:SEQuence:POINt:RISE/FALL
Channel Specific
Command
Description
Sets the rising and falling slew rates for the current
point.
Syntax
:SEQuence:POINt:RISE/FALL<NRf>[A/us]|MIN|MAX
Parameter
<NRf>[A/us], MIN, MAX
Slew rate
Vista de página 150