Gw-instek PEL-3000 Series Programming Manual Manual do Utilizador Página 84

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 174
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 83
PEL-3000 Programming Manual
82
Query example
:RESistance:L1?
> 0
Returns the “level1” conductance setting(MHO).
:RESistance:L2
Set
Query
Description
Sets or queries the CR mode “level2conductance.
This command is applicable to dynamic mode
only.
Note: A different value can be set for each current
range (High/Mid/Low).
Syntax
:RESistance:L2 {<NRf>[ MHO ] | MINimum |
MAXimum }
Query Syntax
:RESistance:L2?
Parameter
<NRf>
level2” conductance
value
MINIMUM or MIN
Minimum conductance
level
MAXIMUM or MAX
Maximum conductance
level
Return parameter
<NR2>
Returns the “Level2
conductance value.
Example
:RESistance:L2 MIN
Sets the conductance value to the minimum.
Query example
:RESistance:L2?
> 2.
Returns the “Level2conductance setting.(MHO)
:RESistance:SET
Set
Query
Description
Sets or queries the CR dynamic mode Set value
when the dynamic units are set to Percent.
This command is only applicable in dynamic
mode.
Vista de página 83
1 2 ... 79 80 81 82 83 84 85 86 87 88 89 ... 173 174

Comentários a estes Manuais

Sem comentários