Gw-instek GSP-930 Programming Manual Manual do Utilizador Página 131

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 195
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 130
REMOTE CONTROL
129
Syntax
[:SENSe]:CNR:CHANnel:SPACe <freq>
Query Syntax
[:SENSe]:CNR:CHANnel:SPACe?
Parameter
<freq>
<NRf>
Return parameter
<NR3>
Hz
Example
:CNR:CHAN:SPAC 6.0e+6
[:SENSe]:CNR:DELTamarker:MODE
Set
Description
Turns the CNR Noise Marking function to
Min(AUTO) or ΔMarker(MANual).
Syntax
[:SENSe]:CNR:DELTamarker:MODE {AUTO|MANual }
Parameter
AUTO
Sets the Noise Marking to Min.
ΔMarker
Sets the Noise Marking to ΔMarker.
Example
:CNR:DELT:MODE AUTO
[:SENSe]:CORRection:CSET<n>:DATA
Set
Query
Description
As a command, sets an offset for a certain
frequency for a selected correction set.
As a query, returns the data contents for the
selected correction set as <csv data>. The data will
be arranged as:
pt#1 freq, pt#1 offset, pt#2 freq, pt#2 offset,...
Syntax
[:SENSe]:CORRection:CSET<n>:DATA <freq>,<offset>
Query syntax
[:SENSe]:CORRection:CSET<n>:DATA?
Parameter
<freq>
<NRf> Hz
<offset>
<NRf> dB
<n>
<NR1>correction set number
Return parameter
<CSV data>
pt#1 freq, pt#1 offset,….. pt#n freq, pt#n
offset
Example
CORR:CSET1:DATA 2e+6,30
Vista de página 130
1 2 ... 126 127 128 129 130 131 132 133 134 135 136 ... 194 195

Comentários a estes Manuais

Sem comentários