REMOTE CONTROL
25
2. Connect a GPIB cable from a
GPIB controller to the GPIB port
on the spectrum analyzer.
3. Turn the spectrum analyzer on.
4. Press
System
>More 1/2[F7]>RmtInterface
Config[F2]>GPIB Addr and set the GPIB
address.
GPIB address
0~30
GPIB constraints
Maximum 15 devices altogether, 20m cable length,
2m between each device
Unique address assigned to each device
At least 2/3 of the devices turned On
No loop or parallel connection
Configure the LAN and LXI Interface
The GSP-930 is a class C LXI compliant instrument. The LXI
specification allows instrumentation to be configured for basic
remote control or monitoring over a LAN.
For details on the LXI specification and compliance classes, please
see the LXI website @ http://www.lxistandard.org.
Background
The LAN interface is used for remote control
over a network. The spectrum analyzer
supports DHCP connections so the instrument
can be automatically connected to an existing
network. Alternatively, network settings can
also be manually configured. The LAN port
supports Auto-MDIX.
LAN
configuration
Settings
IP Address
Default Gateway
Subnet Mask
DNS Server
DHCP on/off