REMOTE CONTROL
141
Syntax
[:SENSe]:PMETer:LLIMit <ampl>
Query Syntax
[:SENSe]:PMETer:LLIMit?
Parameter
<ampl>
<NRf> power unit, default = dBm
Return parameter
<NR3>
Unit = current unit.
Example
:PMET:LLIM 0
[:SENSe]:PMETer:PSENsor:MODE
Set
Query
Description
Sets or queries the power meter sensor mode.
Syntax
[:SENSe]:PMETer:PSENsor:MODE
{LOWNoise|FASTer}
Query Syntax
[:SENSe]:PMETer:PSENsor:MODE?
Parameter/
Return parameter
LOWNoise
Sets the power meter mode to low noise.
FASTer
Set the power meter mode to fast.
Example
:PMET:PSEN:MODE
[:SENSe]:PMETer:RECording:TIME
Set
Query
Description
Sets or queries the power meter recording time.
Syntax
[:SENSe]:PMETer:RECording:TIME
<hour>,<minute>,<second>
Query Syntax
[:SENSe]:PMETer:RECording:TIME?
Parameter/
Return parameter
<hour>
<NR1>Recording time: hours.
<minute>
<NR1>Recording time: minutes.
<second>
<NR1>Recording time: seconds.
Example
:PMET:REC:TIME 1,10,30
[:SENSe]:PMETer:RECording:TIME:STEP
Set
Query
Description
Sets or queries the power meter recording interval
in seconds.
Syntax
[:SENSe]:PMETer:RECording:TIME:STEP <time>